常見問題

    與客服聯絡

    檔案下載

透析

常見問題

請問 - 該如何縮短Dialysis時間?提升效率!

您好:

Parameters to Increase Rate of Dialysis

Sample & Buffer Concentrations

  1. equal except for retained and transferred species

Sample to Dialysate Ratio

  1. 1 : 1000+ by volume

Membrane MWCO

  1. 90-120% of the retained species molecular size

Tubing Length (Surface Area : Volume Ratio)

  1. 30 cm ( 0.1 – 1ml / cm)

Temperature

  1. Elevated (36-60 Celsius)

Dynamic Dialysis

Buffer Changing

  1. 4-5 changes at 2, 4, 6-8, 8-10, 14-16 & 20 hours

Duration (Resulting)

  1. 0.5 – 1.5 days
top